A cell-free sensor platform for the quantification of arsenic concentrations in drinking water.

Notifications You must be signed in to change notification settings

Biomaker/01_A-cell-free-sensor-for-arsenic-in-drinking-water.

Repository files navigation

alt text

Synopsis

Arsenic contamination of drinking water is a global issue affecting in the region of 150 million people. Arsenic poisoning is highly prevalent in countries surrounding the Himalayas such as India, Nepal, Bangladesh and Pakistan and arises from natural contamination of shallow tube wells. Currently, drinking water is predominantly tested via chemical sensors, which require toxic reagents, technical expertise and only produces qualitative results. Developments in synthetic biology have shown that biological components can be engineered for heavy metal detection, but methods are limited by GM contamination risks and only provide qualitative readouts. As such, an opportunity has been identified to develop a first-of -kind cell-free quantitative arsenic sensor which uses the synthesis of an electroactive metabolic enzyme for amperometric quantification of arsenic concentrations. This project was for a first proof of concept of such a sensor.

The system is based on a genetic circuit consisting of a promoter responsive to arsenic and a downstream reporter which produces an electroactive enzyme. The enzyme concentration will then be measured using amperometric detection methods allowing a quantitative measurement of arsenic concentration to be determined.

Sensor Overview

The Figure below summarises the sensor and its development. The sensor is based on two genetic constructs. The first construct provides an inhibiting protein which, when bound to the promoter of the second construct, is sensitive to the presence of arsenic regulating the production of the redox enzyme, glucose oxidase (GOx), by the reporter. The enzyme concentration is measured using amperometric detection, with a potentiostat, to provide a quantitative measurement of arsenic concentration which is then shared to a central database. Thus far the different aspects of the sensor have been developed and tested in isolation and modelling used to consider their interactions.

It is envisaged, that if fully developed the sensor would consist of disposable test strips with printed electrodes and freeze dried cell free extract, genetic circuits and electrochemical reagents. The measurements would be taken via a potentiostat. When the water sample potentially containing arsenic is added, it rehydrates the system allowing amperometric detection.

Synthetic Biology

The Figure shows the interactions of the genetic constructs. (A) Pars arsR construct expresses ArsR protein, (B) ArsR protein inhibits glucose oxidase repressed/activated construct, (C) Arsenic binds to ArsR protein releasing the inhibition and GOx enzyme is expressed.

The repressor Pars arsR plasmid used was from the international Genetically Engineered Machines Competition (iGEM) registry with a pSB1C3 backbone. A number of different variations of the repressed/activated plasmid Pars GOx were developed through cloning. Pars GOx Mvenus was developed for testing with the production. Testing in the presence of arsenic was conducted first in vivo, within ecoli cells, then in TX-TL cell free extract in vitro as would be used for the final sensor. Testing resulted in the genetic constructs producing GOx correlated to sample arsenic concentrations.

Electrochemistry

The arsenic sensor is based on the same principles and electrochemical readout as well-established glucose sensors but applied in reverse to detect GOx concentrations. Such a system was adapted to a glucose oxidase sensor with fixed quantities of glucose and varied glucose oxidase. Different detection methods were trialled, interference tests were then undertaken with cell free extract, iron (III), arsenic and Nepal well water and lastly improvements in terms of limit of detection and sample volume were investigated.

The figure below summarises the different trailed techniques for detection of the GOx enzyme.

Electronics

The figure is a block diagram of the key components of the developed potentiostat. In summary, the required redox potential is set by the user. This is converted into an analogue voltage by signal processing via a digital to analogue converter which is then applied to the potentiostat block. The potentiostat then sets this voltage to the electrochemical cell and can be considered as a single Op amp. The IV converter monitors the currents of the working electrodes and converts them into corresponding voltages by an current-to-voltage converter which is fed into the signal processing in which an analogue to digital converter is used to translate the voltage into a measurable signal corresponding to the cell current.

The figure below shows the user interface developed in Python QT designer. The system is platform independent developed to run on a Raspberry Pi, can be powered by two 9V batteries and has four current ranges of 250nA, 2.5μA, 25μA, 250μA. The developed system can measure currents in the pico amp range with an accuracy of 0.2% of range.

Datasharing

The key objective for Data Sharing was to provide a platform for which measured arsenic readings, taken in rural remote locations, could be transmitted to a central database and displayed online. The system was developed with the consideration of the end users in mind being those who are using the sensor system to test water in countries such as Nepal/Bangladesh and those accessing the data such as NGOs, governments and scientists.

The diagram shows the different aspects of the implemented system. The system required developments in both software and hardware. Both the transmitter and receiver are based on open source Arduino technologies with a Adafruit FONA 808 shield - mini cellular GSM + GPS for Arduino, plug and play shield to provide a GPS location and the ability to transmit the data, by text, over GSM. The system then processes the text in Python to upload it into a SQLite database which is then mapped in Google maps. The system’s capabilities were tested by sending texts with readings from different locations in Cambridge. The cost of the system is currently £130 but is estimated that this could be reduced to approximately £22 if manufactured.

License

The MIT License

Copyright (c) 2010-2017 Google, Inc. http://angularjs.org

Permission is hereby granted, free of charge, to any person obtaining a copy of this software and associated documentation files (the "Software"), to deal in the Software without restriction, including without limitation the rights to use, copy, modify, merge, publish, distribute, sublicense, and/or sell copies of the Software, and to permit persons to whom the Software is furnished to do so, subject to the following conditions:

The above copyright notice and this permission notice shall be included in all copies or substantial portions of the Software.

THE SOFTWARE IS PROVIDED "AS IS", WITHOUT WARRANTY OF ANY KIND, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE AND NONINFRINGEMENT. IN NO EVENT SHALL THE AUTHORS OR COPYRIGHT HOLDERS BE LIABLE FOR ANY CLAIM, DAMAGES OR OTHER LIABILITY, WHETHER IN AN ACTION OF CONTRACT, TORT OR OTHERWISE, ARISING FROM, OUT OF OR IN CONNECTION WITH THE SOFTWARE OR THE USE OR OTHER DEALINGS IN THE SOFTWARE.

About

A cell-free sensor platform for the quantification of arsenic concentrations in drinking water.

Resources

Stars

Watchers

Forks

Releases

No releases published

Packages

No packages published