Automated ultrasound reverberation air image analysis

Automated analysis software and web browser-based results viewer for ultrasound reverberation air images
Documentation for analysis program (currently only in Finnish) »
Documentation for web-based visualization program (currently only in Finnish)»

Report a bug

Contents

  1. About Project
  2. Getting started
  3. Usage
  4. License
  5. Acknowledgements
  6. Contacts

About Project

The goal was to make automated analysis software similar to Horssen et al. 2017 that would allow the user to perform automatic image quality analysis based on ultrasound reverberation images for ultrasonic air images.

The software is divided into two parts:

  1. Automatic analysis of ultrasonic air images
  2. Web-browser for result viewing

Please note: Although the results viewer is browser-based, the software has only been tested in a local-environment in a browser/web server.

Getting started

The software is implemented in Python (v3.8) and the required python library modules are listed in the file py38.yml.

Software download and creation of a new virtual environment

  1. Clone repository:
    git clone https://github.com/MIPT-Oulu/Ultrasound_IQ_analysis.git
  2. Go to root directory, where py38.yml is located:
    cd ..\Ultrasound_IQ_analysis\
  3. Install libraries in your own virtual environment using conda:
    conda env create --file py38.yml
    conda activate py38

.../Ultrasound_IQ_analysis/QA_analysis/data/ directory contains two different ultrasound air image examples that can be used to test the operation of the software.

This software does not support retrieving images directly from the hospital's PACS system. The [Pynetdiccom] (https://pydicom.github.io/pynetdicom/stable/index.html) library can be used to create a DICOM network protocol for an analysis machine.

Retrieval of DICOM images from the PACS system can be done, for example, with the open [k-PACS software] (https://image-systems.biz/products/free-dicom-pacs-tools/k-pacs/).

The user must specify the directory location (in k-PACS software * .. \ imagebox *) where the DICOM files from the quality images arrive and where the watchdog listener path is set. The analysis program listens to the files received in the directory, and in the case of an ultrasound image dicom file, the software performs an automatic quality analysis.

Usage

The use of the software can be divided into two parts: 1. analysis program and 2. results review program.

1. Operation and use of the automatic ultrasound air image analysis program

Operating principle in brief

Running the /QA_analysis/main.py file will start the automatic analysis software (see the next section Usage for more detailed instructions first). It is based on the watchdog listener, which listens to the directory path to which the quality DICOM imagesare sent after PACS query. If a new file appears in the directory path, the analysis of the image is started. Before analysis, check that: 1. the file is of the DICOM type (there is an *.dcm extension) and 2. the image is an ultrasound air image (The patientID is in the correct format as defined in the Settings.yaml file). The image is then analyzed with the same parameters as Horssen et al. 2017 has been described and stored in the result directory based on the following folder hierarchy:

+-- path_save
|   +-- Instituutio_laitevalmistaja_laitteennimi
|        +-- Transducer_1
|             1
|             2 ... etc.
|        +-- Transducer_2
|        +-- Transducer_3 ... etc.
|   +--   Instituutio2_laitevalmistaja2_laitteennimi2
|        +-- Transducer_1
|             1
|             2 ... etc.
|        +-- Transducer_2 ... etc.

In other words, each ultrasound transducer of each ultrasound device has its own folder in which the measurements of a different time point are stored with sequential numbering. The name attributes are automatically retrieved from the image DICOM header data. If the header information is missing, it will be replaced with a generic identifier. However, it is important that each ultrasound device can be directly identified by DICOM header attributes so that the arrangement hierarchy can be implemented automatically. The arrangement hierarchy is utilized in the results viewer.

Instituutio_laitevalmistaja_laitteennimi= Institutional Department Name Attribute + _ + Manufacturer's Model Name Attiribute+ _ + Station Name Attribute
Transducer_1 = TransducerType Attribute + _ + Transducer Frequency Attribute
File e.g. 1 contains python dictionary in which the analysis results has been stored.

Note: It is very typical that dicom header attribute data is partially missing if, for example, if ultrasound images are transferred to an analysis computer via proxy server without RIS.

LUT.xls is a manual look-up table used when the transduver name is not found in the metadata. The name is combined with the Physical deltaX and Physical deltaY parameters (assuming they are found in the metadata if the ultrasound transducer name is missing). More information (currenly only in finnish) documentation.

Usage

Before starting the program:
Go to the Ultrasound_IQ_analysis/QA_analysis directory and open the Settings.yaml file. The user must specify the following information:

  • data_path: the directory location where the DICOM files will be sent to. Note: The Python filename uses a linux notation, i.e. a slash (/) NOT a backslash. The name must also end with a slash, ie ../XX/ZZ/
  • save_path: tiedosto jonne analyysitulokset tallennetaan. HOX: Python tiedostonimi käyttää linux notaatiota eli vinoviivaa (/) EI kenoviivaa.
  • threshold_val: (percentage threshold value, default = 10%), which performs as a threshold for analysis values between previous measurement and current measurements relative difference. I.e. if the relative error is larger than threshold value the value is written in a separate log file (log_file.txt) as additional information for deviating transducers which can be viewed in the web-browser.
  • id_us_analysis: e.g. 112233ULTRA Patient ID identificator, which has to be set same for each air image.
  • path_LUT_table: Directory path for LUT.xls table.

Running the analyssi program from the command line (windows anaconda prompt, linux terminal tms.):

  1. Activate the python virtual environment:
    conda activate py38
    
  2. Go to the directory where the software (main.py) is located:
    cd ...\QA_analysis
    
  3. Start the program:
    python main.py
    
  4. To test the operation of the program Cut + Paste sample files Ultrasound_IQ_analysis/QA_analysis/data to the path_data directory location (Note! If the DICOM images are already in the directory when the program is started, the analysis will no automatically start as the listener is just activated. You need to move the files out of the directory and move back so that the watchdog listener "catches" the files and the analysis is started.

2. Usage of the results web-browser viewer

Operating principle in brief

The purpose of the results review program is to monitor the results of the air image analysis. The viewer was done using Flask microframework and runs in a web browser. Logo

Figure 1. An illustration of the operation of a web browser-based analysis results viewer.

The program is very simple: by selecting Log from the top navigation bar (Figure 1 green) it opens log data from the analyzes; To view the results, select Devices from the navigation bar at the top (Figure 1 blue), which opens the device list. Selecting the device opens a list of ultrasound transducers and selecting the transducer opens a list of measurements and trend monitoring. Logo

Figure 2. An illustration of the operation of a web browser-based analysis results viewer. It has been implemented with Plotly javascript library.

Usage

Before starting the program: Go to the Ultrasound_IQ_analysis/Visualization_app/ directory and open the Settings.yaml file. The user must specify the following information:

  • path_data: The file location where the analysis results are located.
  • path_log: The location where the log file is located(default: .../Ultrasound_IQ_analysis/QA_analysis/log.txt).
  • n_samples: the number of samples (default = 4), which determines how many of the previous quality measurements are moving averaged and moving standard deviated for the trend monitoring. If there exists fewer measurements than defined, the program will take as many measurements as the statistics have been measured.

Running a program from the command line (windows anaconda prompt, linux terminal):

  1. Activate the python virtual environment:
    conda activate py38
    
  2. Go to the directory where the application.../Visualization_app/app.py is located:
    cd ...\Visualization_app
    
  3. Run the program:
    python app.py
    
  4. Open a browser and copy the local http address into the browser. (Note: the program is still under development and therefore has not been set to run through Gunicorn or a similar python web server.)

_Please note: More detailed documentation (only in Finnish) on the operation of the program can be found from .../QA_analysis/documentation and .../Visualization_app/documentation osioista._

License

This project is MIT licensed. See See the LICENSE section for more information.

Acknowledgements

We would like to thank Tampere University Hospital for the ultrasound air images as the test material.

Contacts

The project has been implemented at the University Hospital of Oulu and the Research Unit of Medical, Imaging Physics and Technology at the University of Oulu

The project involves: Satu Inkinen - satu.inkinen [@] oulu.fi
Vili Tuppurainen - vili.tuppurainen [@] ppshp.fi Tuomo Starck - tuomo.starck [@] ppshp.fi
Matti Hanni - matti.hanni [@] ppshp.fi
Miika Nieminen - miika.nieminen [@] ppshp.fi

Project link: https://github.com/MIPT-Oulu/Ultrasound_IQ_analysis


Logo

Automaattinen ultraääni-ilmakuvien analyysi

Ultraääni-ilmakuville automaattinen analyysiohjelmisto ja web-selainpohjainen tulosten tarkasteluohjelma
Dokumentaatio analyysiohjelma »
Dokumentaatio visualisointiohjelma »

Raportoi Bugi

Sisällysluettelo

  1. Tietoa projektista
  2. Alkuasetukset
  3. Käyttö
  4. Lisenssi
  5. Acknowledgements
  6. Yhteystiedot

Tietoa projektista

Tavoitteena oli tehdä Horssen et al. 2017 tutkimuksen mukainen automaattinen analyysiohjelmisto, jonka avulla käyttäjä voi suorittaa ultraääni-ilmakuville automaattisen kuvanlaatuanalyysin reverberaatiokuviin perustuen.

Projektin analyysi jakautuu kahteen osaan:

  1. Ultraääni-ilmakuvien automaattiseen analyysiin
  2. Web-selainpohjaiseen tulosten tarkasteluohjelmaan

Huom: Vaikka tulosten tarkasteluohjelma on selainpohjainen, ohjelmistoa ei kuitenkaan ole testattu kuin paikallisesti selaimessa eikä web-serverikäytössä.

Alkuasetukset

Ohjelmisto on toteutettu Pythonilla (v3.8) ja tiedostosta py38.yml löytyy listattuna on vaadittavat python kirjastomoduulit.

Ohjelmiston lataus ja uuden virtuaaliympäristön luonti

  1. Kloonaa repository:
    git clone https://github.com/MIPT-Oulu/Ultrasound_IQ_analysis.git
  2. Mene root hakemistoon jossa py38.yml sijaitsee:
    cd ..\Ultrasound_IQ_analysis\
  3. Asenna kirjastot omaan virtuaaliympäristöön conda:a käyttäen:
    conda env create --file py38.yml
    conda activate py38

.../Ultrasound_IQ_analysis/QA_analysis/data/ hakemistosta löytyy kaksi erilaista ultraääni-ilmakuva esimerkkikuvaa, joilla ohjelmiston toimintaa voi testata.

Tämä ohjelmisto ei tue kuvien hakemista sairaalan PACS-järjestelmästä suoraan. Pynetdiccom kirjaston avulla voi tehdä DICOM -verkkoprotokollan analyysikoneelle.

Kuvien hakeminen PACS-järjestelmästä voi toteuttaa esimerkiksi avoimella k-PACS ohjelmistolla. Käyttäjän pitää määrittää hakemistosijainti (k-PACS ohjelmistossa ..\imagebox), jonne DICOM tiedostot laatukuvista saapuvat ja johon watchdog kuuntelijan polku asetetaan. Analyysiohjelma kuuntelee hakemistoon saapuneita tiedostoja ja mikäli kyseessä on ultraäänikuva dicom-tietosto, suorittaa ohjelmisto automaattisen laatuanalyysin.

Käyttö

Ohjelmiston käyttö voidaan jakaa kahteen osaan 1. analyysiohjelmaan ja 2. tulosten tarkasteluohjelmaan.

1. Ilmakuvien automaattisen analyysiohjelman toiminta ja käyttö

Toimintaperiaate lyhyesti

/QA_analysis/main.py tiedostoa ajamalla käynnistetään automaattinen analyysiohjelmisto (ks seuraava osio käyttö tarkempi ohjeistus ensin). Se pohjautuu watchdog kuuntelijaan, joka kuuntelee hakemistopolkua, johon laatukuvat tulevat. Mikäli hakemistopolkuun ilmestyy uusi tiedosto aloitetaan kuvalle analyysi. Ennen analyysia tarkistetaan että: 1. tiedosto on DICOM tyyppiä (on *.dcm pääte) ja 2. kuva on ultraääni-ilmakuva (PatientID on oikeaa muotoa, joka määritellään Settings.yaml tiedostossa). Tämän jälkeen kuvasta analysoidaan samat parametrit mitä Horssen et al. 2017 työssä on kuvattu ja ne tallenetaan tuloshakemistoon pohjautuen seuraavanlaiseen kansiohierarkkiaan:

+-- path_save
|   +-- Instituutio_laitevalmistaja_laitteennimi
|        +-- Transducer_1
|             1
|             2 ... etc.
|        +-- Transducer_2
|        +-- Transducer_3 ... etc.
|   +--   Instituutio2_laitevalmistaja2_laitteennimi2
|        +-- Transducer_1
|             1
|             2 ... etc.
|        +-- Transducer_2 ... etc.

Eli siis jokaisen laitteen anturille tulee oma kansionsa johon eri aikapisteen mittaukset tallennetaan juoksevalla numeroinnilla. Nimi attribuutit haetaan automaattisesti kuvan dicom tiedoista. Mikäli tieto puuttuu niin se korvataan jollakin geneerisella tunnisteella. Kuitenkin on tärkeää että jokaisen ultraäänilaitteen voi suoraan tunnistaan dicom attribuuteista, jotta järjestelyhierarkkia voidaan toteuttaa automaattisesti. Järjestelyhierarkkiaa käytetään hyödyksi tulosten katseluohjelmassa.

Instituutio_laitevalmistaja_laitteennimi= Institutional Department Name Attribute + _ + Manufacturer's Model Name Attiribute+ _ + Station Name Attribute
Transducer_1 = TransducerType Attribute + _ + Transducer Frequency Attribute
Tiedostonumero esim. 1 sisältää python dictionaryn, johon analyysitulokset ovat tallennettu.

Huom: On hyvin tyypillistä että dicom tietoja jää uupumaan jos esimerkiksi UÄ-kuvat siirretään analyysikoneelle edustapalvelimen kautta ilman lähetettä. LUT.xls on manuaalinen look-up-taulukko, jota käytetään silloin jos anturin nimeä ei löydy metatiedoista. Nimi yhdistetään Physical deltaX ja Physical deltaY parametrien avulla (oletus että ne löytyvät metatiedoista mikäli anturin nimi puuttuu). Lisätietoja dokumentaatiosta.

Käyttö

Ennen ohjelman käyttöönottoa:
Mene hakemistoon Ultrasound_IQ_analysis/QA_analysis ja avaa Settings.yaml tiedosto.
Käyttäjän pitää määritellä seuraavat tiedot:

  • data_path: tiedostosijanti, johon dicom tiedostot tulevat. HOX: Python tiedostonimi käyttää linux notaatiota eli vinoviivaa (/) EI kenoviivaa. Nimen loppussa pitää myös olla vinoviiva eli ../XX/ZZ/
  • save_path: tiedosto jonne analyysitulokset tallennetaan. HOX: Python tiedostonimi käyttää linux notaatiota eli vinoviivaa (/) EI kenoviivaa.
  • threshold_val: (prosenttiluku 10%), joka on kynnystysarvo mittaustuloksen ja nykyisen mittautuloksen välisen suhteellisen eron välille. Mikäli virhe on suurempi kuin asetettu luku kirjautuu log_file.txt tiedostoon lisätietoja poikkeamasta, joka voi tarkastella myös tarkasteluohjelmassa.
  • id_us_analysis: Esim. 112233ULTRA Patient ID tunniste joka pitää laatukuvaa otettaessa asettaa oikein.
  • path_LUT_table: Hakemistopolku LUT.xls taulukkoon.

Ohjelman ajaminen komentoriviltä (windows anaconda prompt, linux terminal tms.):

  1. Aktivoi python virtuaaliympäristö:
    conda activate py38
    
  2. Mene hakemistoon, jossa sovellus (main.py) sijaitsee:
    cd ...\QA_analysis
    
  3. Käynnistä ohjelma:
    python main.py
    
  4. Ohjelman toimintaa testataksesi Cut+paste esimerkkitiedostot [Ultrasound_IQ_analysis/QA_analysis/data] (https://github.com/MIPT-Oulu/Ultrasound_IQ_analysis/tree/main/QA_analysis/data) path_data sijaintiin (Hox! Jos tiedostot ovat jo kansiossa, kun ohjelma käynnistyy ei analyysi starttaa. Tiedostot pitää siirtää hakemistosta pois ja siirtää takaisin hakemistoon jolloin analyysi käynnistyy).

2. Tulosten tarkasteluohjelman käyttö

Toimintaperiaate lyhyesti

Tulosten tarkasteluohjelman tarkoituksena on monitoroida ilmakuvien laatua. Se on tehty Flask:lla ja toimii web-selaimessa. Logo

Kuva 1. Havainnekuva web-selainpohjaisesta analyysitulosten tarkasteluohjelman toiminnasta.

Ohjelman käyttö on tehty hyvin yksinkertaiseksi, siten että valitaan yläreunan naviogintipalkista Log (kuva 1 vihreä), josta aukeaa lokitiedot analyyseistä. Tulosten tarkasteluun yläreunasta valitaan naviogintipalkista Devices (kuva 1 sininen), joka avaa laitelistauksen (ks kansiohierarkkia). Valitsemalla laitteen avautuu listaus antureista ja anturia painamalla avautuu listaus anturille tehdyistä mittauksista ja trendiseurannasta.

Logo

Kuva 2. Havainnekuva web-selainpohjaisesta analyysitulosten tarkasteluohjelman toiminnasta. Se on toteutettu Plotly javascript kirjastolla.

Käyttö

Ennen ohjelman käyttöönottoa: Mene hakemistoon Ultrasound_IQ_analysis/Visualization_app/ ja avaa Settings.yaml tiedosto.
Käyttäjän pitää määritellä seuraavat tiedot:

  • path_data: Tiedostosijanti jossa analyysitulokset sijaitsevat.
  • path_log: Sijainti jossa loki-tiedosto sijaitsee (oletus: .../Ultrasound_IQ_analysis/QA_analysis/log.txt).
  • n_samples: näytemäärä (4), joka määrittelee kuinka monesta edelliststä laatumittauksesta määritetään keskiarvo ja hajonta-arviot trendiseurantaan. Mikäli mittauksia on vähemmän kuin määritelty niin ohjelma ottaa niin monta mittausta statistiikan laskentaa on tehty.

Ohjelman ajaminen komentoriviltä (windows anaconda prompt, linux terminal):

  1. aktivoi python virtuaaliympäristö:
    conda activate py38
    
  2. Mene hakemistoon, jossa sovellus .../Visualization_app/app.py sijaitsee:
    cd ...\Visualization_app
    
  3. Käynnistä ohjelma:
    python app.py
    
  4. Avaa selain ja kopioi lokaali http osoite selaimeen. (Huom. ohjelmaa on vielä kehitysvaiheessa ja siksi se ei toimi Gunicorn tai vastaavan python verkkopalvelimen kautta.)

_Huom: Tarkempi kuvallinen suomenkielinen dokumentaatio ohjelmien toiminnasta löytyy .../QA_analysis/documentation ja .../Visualization_app/documentation osioista._

Lisenssi

Tämä projekti on MIT lisensoitu. Ks. LICENSE osiosta lisätietoja.

Acknowledgements

Kiitokset Tampereen yliopistolliselle sairaalalle ultraääni-ilmakuva testiaineistosta.

Yhteystiedot

Projekti on toteutettu Oulun yliopistollisessa sairaalassa ja Oulun yliopistossa lääketieteellisen, kuvantamisen fysiikan ja tekniikan tutkimusyksikössä.

Projektissa ovat mukana:
Satu Inkinen - satu.inkinen [@] oulu.fi Vili Tuppurainen - vili.tuppurainen [@] ppshp.fi Tuomo Starck - tuomo.starck [@] ppshp.fi
Matti Hanni - matti.hanni [@] ppshp.fi
Miika Nieminen - miika.nieminen [@] ppshp.fi

Projektin linkki: https://github.com/MIPT-Oulu/Ultrasound_IQ_analysis

About

Ultrasound reverberation pattern analysis for medical image quality asessment.

Resources

Stars

Watchers

Forks

Releases

No releases published

Packages

No packages published